Home » Products » Plasma Protein » Attelica NEPH 630 System
Atellica NEPH 630 System
Smart, simple, and secure protein testing.
Atellica® NEPH 630 System combines proven nephelometric testing with the industry’s broadest menu of plasma protein samples.
- More than 70 assay protocols accommodate multiple sample types.
- Continuous random-access processing provides flexible sample management, easy calibration, and reliable results.
- Wide initial measuring range reduces the need for retesting.
- Intelligent and secure software offers password-protected, predefined user-access profiles and integrated sample result traceability.
Expand testing with the broadest protein assay menu
Atellica NEPH 630 System features the industry’s largest nephelometric menu of plasma protein assays.
Streamline workflows and protect data
Intelligent software supports safe operation and timely data transmission, safeguards patient data, and facilitates quick, convenient technical support, if needed.
Save time and reduce waste with continuous access and smart technology
Continuous random access and a fast average throughput of 65 tests per hour support flexible sample processing and rapid results.
Reduce hands-on time
High-capacity storage and automated features reduce hands-on time.
Deliver accurate and secure results
Proven nephelometric technology offers high precision and reproducibility that provide analytical advantages over turbidimetry in terms of precision and sensitivity.
Empower clinicians to improve patient outcomes
Innovative assays coupled with trusted and proven technologies support clinicians with reliable insight into disease determination and progression.
Allergic Diseases
Anemia/Iron Metabolism
Autoimmune/Rheumatoid Diseases
Blood-CSF Barrier Dysfunction
Cardiovascular Risk/Acute Cardiac Care
Chronic Alcohol Abuse
Coagulation Disorders
Complement Activity
Inflammation
Kidney Disease
Nutritional Assessment
Other Specialty Analytes
Polyclonal and Monoclonal Gammopathies/Immune System
Measuring principle
Nephelometry; measurement of scattered light intensity at a fixed angle of 13–24°
Methods
More than 70 programmed assay protocols
Sample throughput
Effective: approximately 65 tests/hour, depending on the assay mix
Nominal: 100 tests/hourAnalysis method
Fixed-time kinetics, end-point measurement, VLin integral
Calibration
Multi-point calibration
Reagent rotor
Segments for 3 control serum vials
Segments for 2 reagent vialsSample rotor
Segments for 15 sample tubes
Segments for 7 standard vialsDilution unit
1 frame for max. 96 dilution cups
Size of sample tubes
Diameter: 11–16 mm
Height: 65–100 mm
For pediatric samples: Conical microtubes with a maximum filling volume of 1.5 mL
Bar-code reader
Automatic reading of different bar-code types: 2/5 interleaved, Codabar, Code 39, Code 93, Code 128
Test Throughput
Up to 200 tests per hour in batch or random-access mode
Reagent volume
Average consumption: 40 μL per test
Sample dilution
1:1 to 1:32,000
Level detection
For samples, standards, controls, reagents, and container for system liquid and waste
Test Throughput
Up to 200 tests per hour in batch or random-access mode
Reaction cuvettes
Capacity: 90 disposable cuvettes
Measuring temperature
37°C, ±1.5°C
Light source
Infrared high-performance LED
Wavelength
840 nm, ±10 nm
Detector
Photodiode with integrated preamplifier
Connectivity (Sold separately)
Laboratory information system (download mode, host query mode, ASTM)
Smart Remote Services
Atellica Process Manager
Atellica Data Manager
Instrument weight and dimensions
Analyzer dimensions (W x H x D)
107 x 60 (lids closed) x 63 cm (42.1 x 23.6 x 24.8 in.)
Computer and monitor dimensions
(W x H x D including keyboard)
Approx. 36 x 46 x 55 cm (14.25 x 18 x 21.5 in.)
Printer dimensions (W x H x D)
Approx. 41 x 27 x 37 cm (16 x 10.5 x 14.5 in.)
Analyzer weight
115 kg (254 lb)
Computer and monitor weight
Approx. 14 kg (31 lb)
Printer weight
Approx. 7 kg (16.1 lb)
Room environment
Ambient temperature
18–32°C (64–89°F)
Relative humidity
<85% (noncondensing)
Average thermal output
404 W in operating mode
Water requirements
Deionized water NCCLS Type 2
Microbial count must not exceed 100 CFU/mL
Voltage
100 V rating: 90–110 V
120 V rating: 108–132 V
230 V rating: 207–253 V
Power consumption
500 VA (operating mode)
Host interfacing
Download mode, host-query mode, ASTM
Ports
1 10-Base-T/Base TX Ethernet RJ45
1 serial (DB-9) (male)
Why Choose Us
We believe that success is achieved with professionalism, integrity and perfect teamwork, built on the high moral values of our employees.
Our policy is based on impeccable planning, control and execution, through which we achieve maximum satisfaction of our customers.
Professional
customer service
A wide range of laboratory analyzers
Trained and experienced specialists
Certificate
ISO 13485:2012